  • #1175
    Markus Stoeckli
    Moderator

    This error was not reported before. Please check that you have no active processing method. Also, the imaging software needs to run on the actual instrument computer.

    markus

    #1219
    Andy Mahan
    Participant

    Hi – Yes I have seen this error. I am also imaging on a 4700. The only way I found to avoid it, is to not have a processing method open when you start the acquistion. I dont’ believe that there’s sufficient processing ability or time for the 4000 explorer software to keep up during an imaging experiment.

    Andy Mahan