Front Page Forums MMSIT Problems using MMSIT

Viewing 3 posts - 1 through 3 (of 3 total)
  • Author
    Posts
  • #846
    Markus Stoeckli
    Moderator

    This error was not heard of before 🙁 .

    In the software MMSIT, the sample information is stored in a simple MS Access DB. My first guess is, that either your DB got modified or that the shared component for DB interfacing (MDAC) was modified to a different version.

    In a first step, try to run the software with the default DB as it came with the installtion. If that does not help, check on the installation of MDAC on your computer. Use the ‘MDAC Component Checker’ from the MS site [url]http://www.microsoft.com/downloads/details.aspx?familyid=8f0a8df6-4a21-4b43-bf53-14332ef092c9&displaylang=en[/url] to get the relevant information. I just tried it with MDAC 2.7 SP 1 and it works fine. If you have an older version, try upgrading to the most recent one ([url]http://msdn.microsoft.com/data/Default.aspx[/url]). Please report on your Windows version, service packs and MDAC version should the error persist.

    Hope this helps,

    markus

    #877
    wolfrick
    Participant

    Hi Markus!

    I got an error message to reduce the data points/spectra to 32768.
    According to the mass range (2000-25000) and the bin size (4) I have 25108 points /spectra… the error is still present… what to do?

    thanks for your help
    wolfrick

    #878
    Markus Stoeckli
    Moderator

    I checked it out on our sSTR and found nothing wrong with the MMSIT. For the mass range 2000-25000 with bin size 4 ns we end up with 40000 data points in reflector mode and 27000 in linear mode. 💡 Could you verify for me that the data file which you specified in the MMSIT was acquired with the appropriate number of data points (open it as instrument settings in the VC)? It is this file which defines the image settings, no the current settings of the VC.

    Markus

Viewing 3 posts - 1 through 3 (of 3 total)